Introduction to the detection method of common faults of load cells
Sources: | Release date: 2015-11-27 | Browsing volume:
Key words:Introduction to the detection method of common faults of load cells

First of all, before removing the load cell from the weighing system, the structure and sensor of the system should be carefully and carefully determined whether the following problems exist:

(1) Check if the limit device of the system is working and whether the gap meets the requirements;

(2) Check whether the system has damage, rust or obvious wear in the force transmission part; in winter, it should pay attention to whether there is ice formation in the force transmitting part of the sensor, which affects the transmission and reset of the system;

(3) Check if the system is transmitting force failure, which may be due to dust, mechanical parts misalignment, component transmission delay, etc., rather than sensor failure; sensor installation direction is correct.

(4) Check if the sensor cable is connected correctly with the junction box and the display instrument.

Whether there is a bad contact with the disconnected wire;

(5) Check the junction box and the instrument for faults, especially the potentiometer in the junction box.

(6) Check if the sensor is rusted (especially the area of the patch hole); the integrity of the sensor cable; the environment at the entrance of the sensor cable.

Problems with the appearance of the system structure and the sensor are excluded. The following methods can be used for step-by-step testing.

(1) Line rough check

The power supply line, signal line and shielded line of the sensor are coaxial cables, which can be tested with a multimeter (ie power line-signal line, power line-shield line, signal line-shield line), if there is a short circuit, open circuit or A decrease in insulation performance indicates that the sensor may be damaged.

(2) Check the impedance of the load cell

When checking, you need to cut off the power of the electronic scale, open the junction box, and disconnect the signal lines of each sensor. When there is no special detection device, the input impedance Zi and the output impedance Zo of the sensor can be measured with the ohmic file of the 4-digit multimeter, and the measured value is compared with the nominal value on the product certificate provided by the manufacturer. If the side value exceeds the allowable range, it indicates that the sensor may be damaged. (Note: The value of the multimeter used should be accurate, preferably after metering/calibration).


Relevant articles
Related products
  • HT-0720
    HT-0720
  • HT-FWC Stainless steel static load weighing module
    HT-FWC Stainless steel static load weighing module
  • HT-SB Cantilever beam load cell
    HT-SB Cantilever beam load cell
  • FWQ Economic weighing module
    FWQ Economic weighing module
  • HT-GW Column weighing module
    HT-GW Column weighing module
  • HT-3123PTN Weighing control instrument
    HT-3123PTN Weighing control instrument
  • HT-600 Weighing control instrument
    HT-600 Weighing control instrument
  • HT-0745A Stainless steel load cell
    HT-0745A Stainless steel load cell
Contact Us
Add: No. 91, Hanjiang West Road, Xinbei District, Changzhou City
Mob: 0086-519-89619969/0086-519-85505769
Code: 213125
Fax: 0086-519-85505706
E_mail: hoto@dzhq.cn
QR code
Mobile website
WeChat
Submission
Copyright © 2018 Changzhou Hoto Electronic Weighing Apparatus Co., Ltd. All rights reserved Support:Eastnet [Manage]
Top
Baidu
ac米兰app官方下载